Operation

handylab pH 11

34

ba75366e04

01/2011

4.5

Reset

You can reset (initialize) the measurement parameters and the configuration
parameters separately from one another.

Measurement

parameters

The following measured parameters (pH InI) are reset to the default condi-
tion:

Note
The calibration data gets lost when the measuring parameters are reset.
Recalibrate after performing a reset.

Resetting the

measuring

parameters

Measuring mode

pH

Asymmetry

0 mV

Slope -59.16

mV/pH

Calibration procedures

AutoCal TEC

Temperature, manual

25 °C

Resolution (pH display)

0.01 (low resolution)

Display during the
pH calibration

Buffer nominal value

1

Press the <RUN/ENTER> key and hold it down.

2

Press the <CAL> key.

3

Using <

> <>, switch between no and YES.

YES: Resetting the measuring parameters
no: Retaining settings.

4

Confirm with <RUN/ENTER>.
The measuring instrument automatically switches to the pH
measuring mode.