4 configuration, Configuration, Section 4.4 c – Xylem handylab pH 11 User Manual

Page 32: Onfiguration, Ee section 4.4 c

Advertising
background image

Operation

handylab pH 11

32

ba75366e04

01/2011

4.4

Configuration

You can adapt the measuring instrument to your individual requirements. To
do this, the following parameters can be changed (the status on delivery is
marked in bold):

Note
You can leave the configuration menu at any time with <M> . The parameters
that have already been changed are stored.

Display during the

pH calibration

Display during the
pH calibration

Buffer nominal value, or current
electrode voltage

pH resolution

0.01 or 0.001

1

Switch off the measuring instrument.

2

Press the <M> key and hold it down.

3

Press the <ON/OFF> key.
The display test appears briefly on the display. CAL diSP ap-
pears on the display.

4

Select the required display during the pH calibration with
<

> <>.

mV: Display of the current electrode voltage
/pH: Display of the buffer nominal value.

5

Confirm with <RUN/ENTER>.
The measuring instrument automatically switches to the pH
measuring mode.