Xylem handylab pH 11 User Manual

Page 38

Advertising
background image

What to do if...

handylab pH 11

38

ba75366e04

01/2011

No stable mea-

sured value

LoBat display

Buffer solutions

– Incorrect buffer solutions

– Change calibration procedure

– Buffer solutions too old

– Use only once.

Note the shelf life

– Buffer solutions depleted

– Change solutions

Cause

Remedy

pH electrode:

– Diaphragm contaminated

– Clean diaphragm

– Membrane contaminated

– Clean membrane

Test sample:

– pH value not stable

– Measure with air excluded if

necessary

– Temperature not stable

– Adjust temperature if

necessary

Electrode + test sample:

– Conductivity too low

– Use suitable electrode

– Temperature too high

– Use suitable electrode

– Organic liquids

– Use suitable electrode

Cause

Remedy

– Batteries almost empty

– Replace batteries

(see section 5.1 M

AINTENANCE

)