Operation

Multi 3400i

18

4.2.2

Measuring the pH value

AutoRead AR

(Drift control)

The AutoRead function (drift control) checks the stability of the mea-
surement signal. The stability has a considerable impact on the repro-
ducibility of the measured values.

For identical measurement conditions, the following criteria apply:

1

Perform the preparatory activities according to section 4.2.1.

2

Immerse the pH electrode in the test sample.

3

Press the <

> <> keys until pH appears in the status display.

The pH value appears on the display.

4

When measuring without a connected temperature sensor:
Options:
z Determine the current temperature using a thermometer

and, while keeping the <RUN/ENTER> key depressed, en-
ter this temperature value with <

> <>.

z TP display indicator not displayed, socket for the second

probe is free:
Connect the second probe (Oxi or Cond) and immerse it in
the same sample. TP flashes, the temperature is automati-
cally measured using the second probe.

Tref20

Lin

Cal

8

1

S

Sal

O

U

%

mg/l

mV/pH

1/cmSal

AutoCal DIN

Tref25

nLF

AutoCal TEC

RCL

Auto

Store

LoBat

ARng

AR

Time

Baud

Day.Month No.

Ident

Year

mbar

TDS

cm

M

°

F

/

K

%

S/

m

cm

8

4

2

9 9

6

°

C

TP

pH

Reproducibility

Response time

Better than 0.02

> 30 seconds

1

Call up the pH measuring mode with <

> <>.

2

Activate the AutoRead function with <AR>.
The current measured value is frozen (hold function).