4 general errors, General errors – Xylem MULTI 3400 MULTI-PARAMETER METERS User Manual

Page 69

Advertising
background image

Multi 3400i

What to do if...

69

6.4

General errors

Display LoBat

Instrument does not

react

to keystroke

Display

to

to

to

to

Probe symbol flashes

Message

StoFull

StoFull

StoFull

StoFull

Cause

Remedy

– Batteries almost empty

– Exchange the batteries (see

section 5.1 M

AINTENANCE

)

Cause

Remedy

– Operating condition undefined

or EMC load unallowed

– Processor reset:

Hold the <CAL> key
depressed and switch the
instrument on

Cause

Remedy

– Time-out of the interface

– Check that the instrument is

connected

Cause

Remedy

– Calibration interval expired

– Recalibrate the measuring

system

Cause

Remedy

– All memory locations are full

– Output data storage and clear

data storage