9 reset, Reset, Section 4.9 r – Xylem MULTI 3400 MULTI-PARAMETER METERS User Manual

Page 60: Eset

Advertising
background image

Operation

Multi 3400i

60

4.9

Reset

You can reset (initialize) the measurement parameters and the config-
uration parameters separately from one another.

Measurement

parameters

The following measuring parameters (pH/O

2

/Cond InI) can be reset to

the default condition:

pH

O

2

Cond

Note
The calibration data gets lost when the measuring parameters are re-
set. Recalibrate after performing a reset.

Configuration

parameters

The following configuration parameters (InI) are reset to the delivery
status:

Measuring mode

pH

Asymmetry

0 mV

Slope -59.16

mV

Measuring mode

D. O. concentration

Relative slope

1.00

Salinity (value)

0.0

Salinity (function)

Off

Measuring mode

Cell constant

0.475 cm

-1

(fixed)

Baud rate

4800

Display during the
pH calibration

Buffer nominal value

Interval 1
(automatically saved)

OFF

Interval 2
(for data transmission)

OFF