Calibration intervals (Int 3, Int 4, Int 5)

For each measured variable, a time interval is stored. When it has ex-
pired, you will be reminded to calibrate. After a calibration interval has
expired, the probe symbol of the relevant measured variable flashes. It
is still possible to measure. By calibrating the relevant probe, the func-
tion is reset and the interval starts anew.

The following calibration intervals are set in the factory:

Setting the calibration

interval

You can change each of these intervals (1 ... 999 days):

Measured parameter

Designation

Default setting

pH/ORP

Int 3

7 days

Dissolved oxygen

Int 4

14 days

Conductivity

Int 5

180 days

1

Switch off the measuring instrument.

2

Press the <M> key and hold it down.

3

Press the <ON/OFF> key.
The display test appears briefly on the display. After this, the
measuring instrument automatically switches over to configu-
ration.

4

Press the <RUN/ENTER> key repeatedly, until Int 3 ... 5 to-
gether with the required measured variable (pH, O

2

or

) ap-

pears on the display. Example:

5

Set the required interval until the next calibration with <

>

<

>.

6

Confirm with <RUN/ENTER>.

7

Switch to the measuring mode with <M>.

°

C

1

TP

ARng

mg/l

LoBat

Baud

Day.Month No.

Ident

Year

Sal

%

mbar

S

RCL

AR

Store

Auto

°

F

1/cm

/

K

%

TDS

UpH

Sal

mV/pH

S/

m

cm

cm

M

Lin

Tref25 Tref20

nLF

AutoCal DIN

AutoCal TEC

4

1

t 4

n

O

1

d

Time

Cal

Advertising