Maintenance and troubleshooting, Cleaning optical ports, Cleaning fpm 101 optical ports – Extron Electronics FLS 101 User Guide User Manual

Page 23: Repair and calibration, Cleaning optical, Ports

Advertising
background image

Maintenance and

Troubleshooting

This section provides preventative maintenance and troubleshooting suggestions.

Cleaning Optical Ports

Repair and Calibration

Battery Replacement

Mandrels for Multimode Transmit Reference Cables

Loops for Singlemode Single Reference Cables

Cleaning Optical Ports

CAUTION: Before conducting the following procedures be certain the instruments are turned

Off. Disconnect the external supply from the FLS 101.

Optical ports must be kept free from dirt or other contaminants to ensure accurate
measurements and operation. Always clean reference cables before conducting the
test procedures outlined in this guide. It is important to keep dust caps in place when
instruments are not being used.
For cleaning connector end faces on the FLS 101 light sources, test reference cables, and in
fiber frames or adapters, use optical quality cleaning fluid and molded cleaning tips.
For cleaning the FPM 101 optical ports and adapter caps, use lint-free optical cleaning wipes
and optical quality cleaning fluid such as IPA -Reagent Grade Isopropyl Alcohol 99% or
better) and a can of filtered compressed air.

Cleaning FPM 101 Optical Ports

1.

Unscrew the adapter cap from the adapter cap mount.

2.

Use lint-free optical cleaning wipes and optical quality cleaning fluid and dampen a
portion of the wipe with the cleaning fluid.

NOTE: If using isopropyl alcohol (IPA), be sure to use 99% pure IPA that has not

been contaminated.

3.

Gently wipe the exposed FPM 101 port starting with the wet section of the wipe and
pulling it to the dry section.

NOTE: Starting with wet cleaning and finishing with dry improves cleaning action,

reduces static buildup, and leaves the end-face dry.

4.

Using a can of filtered compressed air (held vertically), blow out any contaminants from
the adapter cap.

5.

Once the cleaning is complete, replace the adapter cap.

Fiber Optic Test Meters • Maintenance

17