5 use lucas, Lu c as, 1 arrival at the patient – Physio-Control LUCAS 2 User Manual

Page 15: 2 unpack lucas

Advertising
background image

LUCAS

TM

2 Chest Compression System – Instructions for Use

100666-00 F Valid from CO J2547, © 2011 JOLIFE AB

15

5 Use LUCAS™

5.1 Arrival at the patient

When you have confirmed a cardiac arrest,
immediately start manual cardiopulmonary
resuscitation (CPR). Continue with a
minimum of interruptions.

5.2 Unpack LUCAS™

1. Position the bag with its top nearest to

you.

2. Put your left hand on the black strap on

the left side and pull the red handle so
that the bag unfolds.

3. Push ON/OFF on the User Control Panel

for 1 second to power up LUCAS in the
bag and start the self test. The green
LED adjacent to the ADJUST key illumi-
nates when LUCAS is ready for use.

Note: LUCAS powers down automatically
after 5 minutes if you let it stay in the
ADJUST mode.

Caution - device alarm
If there is a malfunction, the red Alarm LED
illuminates and a buzzer signal is heard. For
trouble shooting, refer to section 8.3.

Caution - keep Battery installed
The Battery must always be installed for
LUCAS to be able to operate, also when
powered by the external Power Supply.

LU

C

AS

2

TM

3

0

:2

3

Advertising