30

9.5 Battery physical specifications

9.6 Battery environmental specifications

Category

Specifications

Size (H × W × D)

5.1 x 3.5 x 2.2 inches / 13.0 Ч 8.8 Ч 5.7 cm

Weight

1.3 lbs / 0.6 kg

Type

Rechargeable Lithium Ion Polymer (LiPo)

Capacity

3300 mAh (typical), 86 Wh

Battery voltage (nominal)

25.9 V

Initial Battery runtime (nominal patient)

45 minutes (typical)

Maximum Battery charge time

Less than 4 hours at room temperature
(72ºF / 22ºC)

Required interval for replacement of the Battery

Recommendation to replace the Battery every 3 years or
after 200 uses (of more than 10 minutes use each time)

Category

Specifications

Operating temperature

32ºF to 104ºF / 0ºC to +40ºC, ambient when installed in the
device

Charge temperature

41ºF to 95ºF / 5ºC to +35ºC ambient (68ºF to 77ºF / 20ºC to
25ºC preferred)

Storage temperature

32ºF to 104ºF / 0ºC to +40ºC ambient for less than six
months

IP classification (IEC60529)

IP 44

Advertising