2 Chest Compression System – Instructions for Use

100666-00 F Valid from CO J2547, © 2011 JOLIFE AB

28

8.3 Malfunction alarms

Below is a list of all alarms that can occur on LUCAS. You mute all alarms for 60 seconds if you
push MUTE.

Start with manual compressions immediately if LUCAS does not operate properly.

If the malfunction described above seems permanent, LUCAS must be examined by approved
service personnel. Please consult your local Physio-Control representative or JOLIFE AB.
Contact information is available at www.JOLIFE.com

Reason

Visual LED indication

Audible alarms

Result

Compression pattern outside
limit (too deep, too shallow and
timing failure)

Red alarm LED

Alarm

Compressions stop

Rising temperature in LUCAS

Warning alarm

None

Too high temperature in LUCAS

Red alarm LED

Alarm

Compressions stop

Hardware error

Red alarm LED

Alarm

Compressions stop

Too high Battery temperature

Intermittent red Battery
warning:
The red Battery indication
LED farthest to the right
illuminates intermittently.

Intermittent alarm

Compressions stop

Battery charge too low

Intermittent red Battery
warning:
The red Battery indication
LED farthest to the right
illuminates intermittently.

Intermittent alarm

Compressions stop.
The Battery must be
recharged in the
external Battery
Charger.

Battery error

Constant red Battery
warning:
The red Battery indication
LED farthest to the right
shows a constant light.

Alarm

Compressions stop.
The Battery cannot be
used anymore.