5

Agilent 8722ES with 2.4 mm test ports

Standard, Options 400, 012

1

, 089, or any combina-

tion of these options.

Calibration kit: 85056A, 2.4-mm with sliding loads
Cables: 85133F 2.4-mm flexible cable set
IF bandwidth: 10 Hz
Averaging: None (8 during isolation calibration)

System dynamic range

0.05 to 0.84 GHz

67 dB

0.84 to 8 GHz

93 dB

8 to 20 GHz

91 dB

20 to 40 GHz

80 dB

2

Magnitude

Phase

Magnitude

Phase

Corrected measurement port specifications

Frequency range (GHz)

0.05 to 2

2 to 8

8 to 20

20 to 40

Directivity

42 dB

42 dB

42 dB

38 dB

Source match

42 dB

38 dB

38 dB

33 dB

Load match

42 dB

42 dB

42 dB

38 dB

Reflection tracking

±(0.005 dB + 0.03 dB/°C)

±(0.010 dB + 0.03 dB/°C)

±(0.010 dB + 0.04 dB/°C)

±(0.021 dB + 0.06 dB/°C)

Transmission tracking

±(0.020 dB + 0.03 dB/°C)

±(0.038 dB + 0.03 dB/°C)

±(0.048 dB + 0.04 dB/°C)

±(0.110 dB + 0.06 dB/°C)

Maximum output power
0.05 to 20 GHz: –5 dBm
20 to 40 GHz: –10 dBm

1. Internal test set not bypassed.
2. 3 dB less with Option 012.