System performance, continued – Atec Agilent-8722ES User Manual

Page 6

Advertising
background image

10

Agilent 8719ES or 8720ES with Option 007 or 085

Calibration kit: 85052B 3.5-mm with sliding loads
Cables: 85131F 3.5-mm flexible cable set
IF bandwidth: 10 Hz
Averaging: None (8 during isolation calibration)

System performance, continued

Corrected measurement port specifications

Frequency range (GHz)

0.05 to 0.5

0.5 to 2

2 to 8

8 to 20

Directivity

48 dB

48 dB

44 dB

44 dB

Source match

40 dB

39 dB

32 dB

30 dB

Load match

48 dB

48 dB

44 dB

44 dB

Reflection tracking

±(0.006 dB + .02 dB/°C)

±(0.010 dB + .03 dB/°C)

±(0.030 dB + .03 dB/°C)

±(0.031 dB + .04 dB/°C)

Transmission tracking

±(0.011 dB + .02 dB/°C)

±(0.016 dB + .03 dB/°C)

±(0.066 dB + .03 dB/°C)

±(0.108 dB + .04 dB/°C)

Magnitude

Phase

Magnitude

Phase

Measurement uncertainty

Reflection measurements

Transmission measurements

Advertising