System performance, continued – Atec Agilent-8722ES User Manual

Page 4

Advertising
background image

8

Agilent 8719ES Option 400, 8720ES Option 400,
and 8722ES Option 400 with 3.5 mm test ports
using TRL

Includes instruments with Options 012

1

and/or 089.

Calibration kit: 85052C, 3.5-mm for TRL
Cables: 85131F 3.5-mm flexible cable set
IF bandwidth: 10 Hz
Averaging: None (8 during isolation calibration)

System dynamic range

Frequency range

8719ES/8720ES

8722ES

0.05 to 0.084 GHz

77 dB

67 dB

0.084 to 8 GHz

100 dB

93 dB

8 to 20 GHz

100 dB

91 dB

20 to 40 GHz

80 dB

2

Maximum output power
8719ES/8720ES: +5 dBm
8722ES (0.05 to 20 GHz): –5 dBm

(20 to 40 GHz): –10 dBm

System performance, continued

Corrected measurement port specifications

8719ES/8720ES

Frequency range (GHz)

Option 400

0.05 to 0.5

0.5 to 2

2 to 8

8 to 20

Directivity

48 dB

48 dB

50 dB

50 dB

Source match

40 dB

40 dB

50 dB

50 dB

Load match

48 dB

48 dB

50 dB

50 dB

Reflection tracking

±(0.006 dB + 0.02 dB/°C)

±(0.006 dB + 0.03 dB/°C)

±(0.005 dB + 0.03 dB/°C)

±(0.005 dB + 0.04 dB/°C)

Transmission tracking

±(0.013 dB + 0.02 dB/°C)

±(0.017 dB + 0.03 dB/°C)

±(0.06 dB + 0.03 dB/°C)

±(0.019 dB + 0.04 dB/°C)

8722ES Option 400

Frequency range (GHz)

0.05 to 2

2 to 8

8 to 20

20 to 26.5

Directivity

48 dB

50 dB

50 dB

50 dB

Source match

40 dB

50 dB

50 dB

50 dB

Load match

48 dB

50 dB

50 dB

50 dB

Reflection tracking

±(0.006 dB + 0.03 dB/°C)

±(0.005 dB + 0.03 dB/°C)

±(0.005 dB + 0.04 dB/°C)

±(0.005 dB + 0.06 dB/°C)

Transmission tracking

±(0.017 dB + 0.03 dB/°C)

±(0.013 dB + 0.03 dB/°C)

±(0.016 dB + 0.04 dB/°C)

±(0.023 dB + 0.06 dB/°C)

Magnitude

Phase

Magnitude

Phase

Measurement uncertainty

Reflection measurements

Transmission measurements

1. Internal test set not bypassed.
2. 3 dB less with Option 012.

Advertising