13

Agilent 8722ES with Option 007 or 085

Calibration kit: 85056D 2.4-mm with fixed loads
Cables: 85133F 2.4-mm flexible cable set
IF bandwidth: 10 Hz
Averaging: None (8 during isolation calibration)

Corrected measurement port specifications

Frequency range (GHz)

0.05 to 2

2 to 8

8 to 20

20 to 40

Directivity

42 dB

42 dB

34 dB

26 dB

Source match

39 dB

36 dB

29 dB

23 dB

Load match

41 dB

38 dB

33 dB

26 dB

Reflection tracking

±(0.011 dB + .03 dB/°C)

±(0.046 dB + .03 dB/°C)

±(0.048 dB + .04 dB/°C)

±(0.090 dB + .06 dB/°C)

Transmission tracking

±(0.022 dB + .03 dB/°C)

±(0.053 dB + .03 dB/°C)

±(0.130 dB + .04 dB/°C)

±(0.367 dB + .06 dB/°C)

Magnitude

Phase

Magnitude

Phase

Measurement uncertainty

Reflection measurements

Transmission measurements