Test port input (dynamic accuracy) – Atec Agilent-E5071C User Manual

Page 25

Advertising
background image

25

Test port input (dynamic accuracy)

1

(continued)

Table 25.

Option 2D5/2K5/4D5/4K5

Accuracy of the test port input power reading is relative to –10 dBm reference input
power level.

Description

Specifi cation

Typical

Dynamic accuracy magnitude

10 dBm

±2.49 dB

–30 dBm

±0.046 dB

–100 dBm

±2.00 dB

–110 dBm

±3.0 dB

Dynamic accuracy phase

10 dBm

±20.6 deg

–30 dBm

±0.30 deg

–100 dBm

±15.0 deg

Specification

±3.0 dB (at –110 dBm, Ref.= –10 dBm, typical)

1. Dynamic accuracy is verifi ed with the following measurements:

• compression over frequency

• IF linearity at two frequencies (1 MHz and 1.195 GHz) using a reference level of –10 dBm for an

input power range of 0 to –100 dBm.

Test port power (dBm)

Accuracy (dB)

Magnitude

0.01

0.1

1

10

-100

-90

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Test port power (dBm)

Accuracy (dB)

Phase

0.1

1

10

100

-100

-90

-80

-70

-60

-50

-40

-30

-20

-10

0

10