Test port input (dynamic accuracy) – Atec Agilent-E5071C User Manual
Page 25
25
Test port input (dynamic accuracy)
1
(continued)
Table 25.
Option 2D5/2K5/4D5/4K5
Accuracy of the test port input power reading is relative to –10 dBm reference input
power level.
Description
Specifi cation
Typical
Dynamic accuracy magnitude
10 dBm
±2.49 dB
–30 dBm
±0.046 dB
–100 dBm
±2.00 dB
–110 dBm
±3.0 dB
Dynamic accuracy phase
10 dBm
±20.6 deg
–30 dBm
±0.30 deg
–100 dBm
±15.0 deg
Specification
±3.0 dB (at –110 dBm, Ref.= –10 dBm, typical)
1. Dynamic accuracy is verifi ed with the following measurements:
• compression over frequency
• IF linearity at two frequencies (1 MHz and 1.195 GHz) using a reference level of –10 dBm for an
input power range of 0 to –100 dBm.
Test port power (dBm)
Accuracy (dB)
Magnitude
0.01
0.1
1
10
-100
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
Test port power (dBm)
Accuracy (dB)
Phase
0.1
1
10
100
-100
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10