Cycle time (ms), Vs. number of points – Atec Agilent-E5071C User Manual

Page 49

Advertising
background image

49

Cycle time (ms)

1, 2

vs. number of points

Table 48.

Option 230/235/240/245/260/265/280/285/430/435/440/445/

460/465/480/485

Start 1 GHz, stop 1.2 GHz, 500 kHz IF bandwidth, error correction: OFF,
display update: OFF, number of traces = 1

Number of
points

Sweep mode:
Fast swept,
system error
correction: OFF

Sweep mode:
Fast swept,
system error
correction: ON

Sweep mode:
Std stepped,
system error
correction: ON

3

3.1

3.1

3.4

11

3.3

3.3

3.3

51

3.5

3.5

4.4

101

3.9

4.1

5.6

201

4.9

5.5

7.7

401

7.1

8.0

12

801

12

14

18

1601

20

24

29

Cycle time (ms)

1, 2

vs. number of points

Table 49.

Option 2D5/2K5/4D5/4K5

Start 11 GHz, stop 12 GHz, 500 kHz IF bandwidth, error correction: OFF,
display update: OFF, number of traces = 1

Number of
points

Sweep mode:
Fast swept,
system error
correction: OFF

Sweep mode:
Fast swept,
system error
correction: ON

Sweep mode:
Std stepped,
system error
correction: ON

51

3.0

3.0

2.8

201

3.2

3.4

6.5

401

4.9

5.9

11

1601

17

20

31

1. Supplemental performance data.