Cycle time (ms)
1, 2
vs. number of points
Table 48.
Option 230/235/240/245/260/265/280/285/430/435/440/445/
460/465/480/485
Start 1 GHz, stop 1.2 GHz, 500 kHz IF bandwidth, error correction: OFF,
display update: OFF, number of traces = 1
Number of
points
Sweep mode:
Fast swept,
system error
correction: OFF
Sweep mode:
Fast swept,
system error
correction: ON
Sweep mode:
Std stepped,
system error
correction: ON
3
3.1
3.1
3.4
11
3.3
3.3
3.3
51
3.5
3.5
4.4
101
3.9
4.1
5.6
201
4.9
5.5
7.7
401
7.1
8.0
12
801
12
14
18
1601
20
24
29
Cycle time (ms)
1, 2
vs. number of points
Table 49.
Option 2D5/2K5/4D5/4K5
Start 11 GHz, stop 12 GHz, 500 kHz IF bandwidth, error correction: OFF,
display update: OFF, number of traces = 1
Number of
points
Sweep mode:
Fast swept,
system error
correction: OFF
Sweep mode:
Fast swept,
system error
correction: ON
Sweep mode:
Std stepped,
system error
correction: ON
51
3.0
3.0
2.8
201
3.2
3.4
6.5
401
4.9
5.9
11
1601
17
20
31
1. Supplemental performance data.
2. Measured with fi rmware version A.09.30.