Measurement throughput summary, Cycle time for measurement completion – Atec Agilent-E5071C User Manual

Page 37

Advertising
background image

37

Measurement Throughput Summary

Cycle time for measurement completion

1, 2

(ms)

Table 36.

Option 230/235/240/245/260/265/280/285/430/435/440/445/

460/465/480/485

Sweep mode: Swept, analyzer display turned off with: DISP:ENAB OFF,
number of traces = 1, system error correction: OFF

Number of points

51

201

401

1601

Start 1 GHz, stop 1.2 GHz, 500 kHz IF bandwidth

Uncorrected

3.4

4.9

7.1

20

2-port cal

6.3

9.3

14

41

4-port cal

13

19

28

92

Start 1 GHz, stop 1.2 GHz, 100 kHz IF bandwidth

Uncorrected

3.7

5.2

7.3

20

2-port cal

6.6

10

15

42

4-port cal

13

21

32

94

Start 1 GHz, stop 1.2 GHz, 1 kHz IF bandwidth

Uncorrected

54

202

397

1563

2-port cal

107

401

792

3125

4-port cal

213

800

1581

6247

Start 100 kHz, stop 3 GHz, 500 kHz IF bandwidth

Uncorrected

8.7

11

12

25

2-port cal

17

21

24

48

4-port cal

34

42

49

102

Start 100 kHz, stop 3 GHz, 100 kHz IF bandwidth

Uncorrected

8.8

11

13

26

2-port cal

17

22

24

54

4-port cal

34

42

49

118

Start 100 kHz, stop 3 GHz, 1 kHz IF bandwidth

Uncorrected

56

206

404

1583

2-port cal

112

410

805

3163

4-port cal

222

818

1608

6324

1. Supplemental performance data.
2. Measured with fi rmware version A.09.30.