Cycle time for measurement completion – Atec Agilent-E5071C User Manual
Page 39
39
Cycle time for measurement completion
1, 2
(ms)
Table 38.
Option 230/235/240/245/260/265/280/285/430/435/440/445/
460/465/480/485
Sweep mode: Swept, analyzer display turned off with: DISP:ENAB OFF,
number of traces = 1, system error correction: ON
Number of Points
51
201
401
1601
Start 1 GHz, stop 1.2 GHz, 500 kHz IF bandwidth
Uncorrected
3.5
5.5
8.0
24
2-port cal
6.4
9.8
15
49
4-port cal
13
20
29
108
Start 1 GHz, stop 1.2 GHz, 100 kHz IF bandwidth
Uncorrected
3.6
5.4
8.0
23
2-port cal
6.6
11
16
48
4-port cal
13
22
33
107
Start 1 GHz, stop 1.2 GHz, 1 kHz IF bandwidth
Uncorrected
54
202
397
1563
2-port cal
107
401
792
3125
4-port cal
213
800
1581
6247
Start 100 kHz, stop 3 GHz, 500 kHz IF bandwidth
Uncorrected
8.7
11
13
28
2-port cal
18
21
24
53
4-port cal
34
42
49
112
Start 100 kHz, stop 3 GHz, 100 kHz IF bandwidth
Uncorrected
8.8
11
13
28
2-port cal
17
22
25
57
4-port cal
34
42
50
122
Start 100 kHz, stop 3 GHz, 1 kHz IF bandwidth
Uncorrected
56
206
404
1583
2-port cal
112
410
805
3163
4-port cal
222
818
1608
6324
1. Supplemental performance data.
2. Measured with fi rmware version A.09.30.