E-Series E9300 Average Power Sensor Specifications

The E-Series E9300 wide dynamic
range, average power sensors are
designed for use with the EPM family
of power meters. These specifications
are valid ONLY after proper calibration
of the power meter and apply for CW
signals unless otherwise stated.

Specifications apply over the tem-
perature range 0 °C to 55 °C unless
otherwise stated, and specifications
quoted over the temperature range
25 °C ±10 °C, conform to the standard
environmental test conditions as
defined in TIA/EIA/ IS-97-A and
TIA/EIA/IS-98-A.

Table 7. E9300 Series two-path specification

“A” suffix sensors

“B” suffix sensors

“H” suffix sensors

High power path

–10 to +20 dBm

+20 to +44 dBm

0 to +30 dBm

Low power path

–60 to –10 dBm

–30 to +20 dBm

–50 to 0 dBm

Table 8. E9300 Series sensors specification

Model

Frequency
range

Maximum SWR
(25 °C ± 10 °C)

Maximum SWR
(0 to 55 °C)

Maximum power

Connector
type

–60 dBm to +20 dBm wide dynamic range sensors

E9300A

10 MHz to 18 GHz

10 MHz to 30 MHz: 1.15
30 MHz to 2 GHz: 1.13
2 GHz to 14 GHz: 1.19
14 GHz to 16 GHz: 1.22
16 GHz to 18 GHz: 1.26

10 MHz to 30 MHz: 1.21
30 MHz to 2 GHz: 1.15
2 GHz to 14 GHz: 1.20
14 GHz to 16 GHz: 1.23
16 GHz to 18 GHz: 1.27

+25 dBm (320 mW) average;
+33 dBm peak (2 W)
(< 10 μsec)

Type-N (m)

E9301A

10 MHz to 6 GHz

10 MHz to 30 GHz: 1.15
30 MHz to 2 GHz: 1.13
2 GHz to 6 GHz: 1.19

10 MHz to 30 MHz: 1.21
30 MHz to 2 GHz: 1.15
2 GHz to 6 GHz: 1.20

+25 dBm (320 mW) average;
+33 dBm peak (2 W)
(< 10 μsec)

Type-N (m)

E9304A

9 kHz to 6 GHz

9 kHz to 2 GHz: 1.13
2 GHz to 6 GHz: 1.19

9 kHz to 2 GHz: 1.15
2 GHz to 6 GHz: 1.20

+25 dBm (320 mW) average;
+33 dBm peak (2 W)
(< 10 μsec)

Type-N (m)

–30 dBm to +44 dBm wide dynamic range sensors

E9300B

10 MHz to 18 GHz

10 MHz to 8 GHz: 1.12
8 GHz to 12.4 GHz: 1.17
12.4 GHz to 18 GHz: 1.24

10 MHz to 8 GHz: 1.14
8 GHz to 12.4 GHz: 1.18
12.4 GHz to 18 GHz: 1.25

0 to 35 °C: 30 W avg
35 to 55 °C: 25 W avg
< 6 GHz: 500 W pk
> 6 GHz: 125 W pk
500 W.μS per pulse

Type-N (m)

E9301B

10 MHz to 6 GHz

10 MHz to 6 GHz: 1.12

10 MHz to 6 GHz: 1.14

0 to 35 °C: 30 W avg
35 to 55 °C: 25 W avg
< 6 GHz: 500 W pk
> 6 GHz: 125 W pk
500 W.μS per pulse

Type-N (m)

–50 dBm to +30 dBm wide dynamic range sensors

E9300H

10 MHz to 18 GHz

10 MHz to 8 GHz: 1.15
8 GHz to 12.4 GHz: 1.25
12.4 GHz to 18 GHz: 1.28

10 MHz to 8 GHz: 1.17
8 GHz to 12.4 GHz: 1.26
12.4 GHz to 18 GHz: 1.29

3.16 W avg
100 W pk
100 W.μS per pulse

Type-N (m)

E9301H

10 MHz to 6 GHz

10 MHz to 6 GHz: 1.15

10 MHz to 6 GHz: 1.17

3.16 W avg
100 W pk
100 W.μS per pulse

Type-N (m)

The E-Series E9300 power sensors
have two independent measurement
paths (high and low power paths) as
shown in Table 7.