848xD Series Diode and 8483A Thermocouple Power Sensor Specifications

Calibration factor uncertainties

These thermocouple and diode power sensors provide extraordinary accuracy,
stability, and SWR over a wide range of frequencies (100 kHz to 110 GHz) and
power levels (–70 dBm to +20 dBm).

Table 13. Typical root sum of squares (rss) uncertainty on the calibration factor data printed on the power sensor

Frequency (GHz)

8483A

8481D

8485D

8487D

R8486D

Q8486D

0.0001

1.3

0.0003

1.2

0.001

1.1

0.003

1.2

0.01

1.2

0.03

1.2

0.05

1.2

0.1

1.2

0.3

1.2

1

1.2

0.8

1.4

1.3

2

1.2

0.8

1.4

1.3

4

0.8

1.7

1.4

6

0.9

1.7

1.4

8

1.0

1.7

1.4

10

1.1

1.9

1.5

12

1.2

1.9

1.5

14

1.1

2.0

1.6

16

1.5

2.1

1.7

18

1.7

2.2

1.7

22

2.7

1.9

26.5

2.8

2.2

3.0

28

2.9*

2.3

3.2

30

3.2*

2.4

3.0

33

3.3*

2.6

3.0

4.2

34.5

2.6

3.0

4.2

37

2.7

3.0

4.2

40

3.0

4.2

42

3.2

4.9

44

2.5

5.1

46

3.8

5.5

48

3.8

5.8

50

5.0

6.2

* These uncertainties only apply to Option 033.
** The 8480 Series sensors in the table do not include discontinued models.