E-Series E9300 Average Power Sensor Specifications

(continued)

Power linearity*

Table 9. E9300 Series power linearity (after zero and cal at ambient environmental conditions) sensor

Sensor

Power

Linearity (25 °C ±10 °C)

Linearity (0 °C to 55 °C)

E9300A, E9301A, E9304A

–60 to –10 dBm
–10 to 0 dBm
0 to +20 dBm

±3.0%
±2.5%
±2.0%

±3.5%
±3.0%
±2.5%

E9300B, E9301B

–30 to +20 dBm
+20 to +30 dBm
+30 to +44 dBm

±3.5%
±3.0%
±2.5%

±4.0%
±3.5%
±3.0%

E9300H, E9301H

–50 to 0 dBm
0 to +10 dBm
+10 to +30 dBm

±4.0%
±3.5%
±3.0%

±5.0%
±4.0%
±3.5%

* After zero and calibration at ambient environmental conditions

Typical E9300A/01A/04A power linearity at 25 °C, after zero and
calibration, with associated measurement uncertainty

Power range

Measurement uncertainty

–30 to –20 dBm

±0.9%

–20 to –10 dBm

±0.8%

–10 to 0 dBm

±0.65%

0 to +10 dBm

±0.55%

+10 to +20 dBm

±0.45%

0.5
0.4
0.3
0.2
0.1

0

–0.1
–0.2
–0.3
–0.4
–0.5

% Error

18

14

10

6

2

–2

–8

–10

–14

–18

–22

–26

–30

20

Power (dBm)

1

0.8
0.6
0.4
0.2

0

–0.2
–0.4
–0.6
–0.8

–1

% Error

–10

–5

0

5

10

15

20

25

30

Power (dBm)

1

0.8
0.6
0.4
0.2

0

–0.2
–0.4
–0.6
–0.8

–1

% Error

–10

–5

0

5

10

15

20

25

30

Power (dBm)

Typical E9300B/01B power linearity at 25 °C, after zero and
calibration, with associated measurement uncertainty

Power range

Measurement uncertainty

–6 to 0 dBm

± 0.65%

0 to +10 dBm

± 0.55%

+10 to +20 dBm

± 0.45%

+20 to +26 dBm

± 0.31%

Typical E9300H/01H power linearity at 25 °C, after zero and
calibration, with associated measurement uncertainty

Power range

Measurement uncertainty

–26 to –20 dBm

± 0.9%

–20 to –10 dBm

± 0.8%

–10 to 0 dBm

± 0.65%

0 to +10 dBm

± 0.55%

+10 to +20 dBm

± 0.45%

+20 to +26 dBm

± 0.31%