Check the trigger mode on the trigger interface whether or

not it’s in “normal” or “single”, and check the trigger level

whether or not it is over the wave range. If yes, please put

the trigger level to the middle position or set the trigger

mode to the “Auto” position. Another way is to choose the

“Auto” button to set up automatically.

6. After the Acquisition is set to Averages or Display Persistence

time is set too long, the waveform refreshes slowly.

It is normal in these settings.

7. The signal is displayed as a ladder-like waveform

1. This phenomenon is normal. The time base may be too

slow. Turn the horizontal SCALE knob to increase

horizontal resolution to improve the display.

2. Maybe the display Type is set to “Vectors”. You could

set it to Dots mode to improve the display.