3-2. Operating Instructions

a. Connect the unit up to the AC power.

b. Turn on the unit by pushing in the POWER on-off switch.

c. Set the FUNCTION indicator position to FREQ A and the gate TIME

indicator to the 1 Sec, position.

CAUTION

1. APPLICATION OF INPUT VOLTAGES HIGHER THAN THE LIMITS LISTED IN

THE SPECIFICATIONS SECTION MAY DAMAGE THE COUNTER. BEFORE

APPLYING ANY SIGNAL TO THE INPUTS, MAKE CERTAIN THAT IT DOES NOT

EXCEED THESE SPECIFIED MAXIMUMS.

2. FREQUENCY COUNTER GROUND POINTS ARE CONNECTED DIRECTLY TO

EARTH GROUND. CONNECT GROUND OF THE CIRCUIT UNDER TEST TO

THE FREQUENCY COUNTER GROUND ONLY

3-3. Frequency Measurements

A. INPUT A (0.1 Hz to 100 MHz):

a. Apply the signal to be measured at input A.

b. Set the FUNCTION indicator to FREQ. A

c. Select the resolution using the GATE TIME selector switch.

d. The frequency is indicated on the display. The gate indicator is lit while

the measurement is in progress, and the display is updated at the end of

each measurement interval.

e.

Engaging the HOLD switch “freezes” the display at the existing reading.

When HOLD is released, the display is updated and resumes counting.

f.

Engage the ATTENUATOR if necessary. When set to x10 (pushed in),

the signal at input A is attenuated by a factor of approximately 10 before it

is applied to the counter. This helps prevent miscounting caused by noisy

or improperly terminated high amplitude signals.

g. Engage the LOW PASS FILTER (LPF) if necessary. When this switch is