B&K Precision 1856D - Manual User Manual

Page 20

Advertising
background image

20

standing waves and cable shunt capacitance.

Waves propagating along a transmission line will be reflected at its ends if the

line is not terminated in its characteristic impedance. These reflections will set up

standing waves in the line, which may cause damage to the signal source or

produce inaccurate measurements. This effect increases as the cable length

reaches one-fourth of the wavelength for the frequency being measured. Standing

waves can be minimized by keeping cable lengths short, or eliminated by providing

a proper termination. RG58 cable, for example, has a characteristic impedance of

50

. The proper termination for this cable would, therefore, be a 50

resistor

between the center conductor and the shield. A standard BNC, 50

cable

termination is most often used for this.

Input A has an input impedance of 1 M

Ω shunted by < 40 pF. Radio frequency

measurements made at this input will experience standing waves and attenuation

due to the shunt capacitance of the cable which adds to the shunt capacitance at

the input. For RF measurements at input A, it is recommended that a 50

termination be used, or that the cable be no longer than three feet (91 cm).

Input C has a built in impedance of 50

, thus the need for an external

termination is eliminated.

In situations where a 50

termination could affect bias voltages or other DC

voltages in the circuit under test, a blocking capacitor should be used.

5-2. Attenuator Probes

When measuring signals at input A that will exceed its maximum input voltage

level (See FIG. 1), a 10:1 attenuating probe such as an oscilloscope probe should

be used.

NOTE

Advertising