Page 17

Advertising
background image

17

3-5. Total Measurements

The totalize mode is used to count the total number of events occurring during a

specific time period. The maximum frequency is 30 MHz.

a. With the FUNCTION switch, select the TOTAL A mode. Gate and unit

settings are ignored.

b. Apply the signal to be measured at input A. The accumulated counts will

be displayed continuously to a maximum count of 999999999. If this is

exceeded, the overflow message, OF, will be displayed.

c. The ATTENUATOR, LPF, and COUPLING switches operate the same as

they do in the frequency measurements modes.

NOTE

THE HOLD SWITCH MAY BE USED TO LATCH THE DISPLAY. THE

COUNTER,

HOWEVER, WILL CONTINUE TO INCREMENT. WHEN THE HOLD SWITCH IS

RELEASED, THE UPDATED COUNT WILL BE DISPLAYED.

3-6. RPM (Revolutions Per Minute) Measurement

a. Apply the signal to be measured at input A. The count will be

displayed as revolutions per minute (RPM). Maximum count is 999999999.

If this is exceeded, the overflow message, OF, will be displayed.

b. The ATTENUATOR, LPF, and COUPLING switches operate the same as

they do in the frequency measurements modes.

3-7. Use of RS-232C Serial Interfacing

1) Hardware/Software Requirements

IBM PC/XT/AT or compatible computer

Microsoft Windows

Serial port for connection with counter

2) Output Data Formats