3-6

Section 3 Operation

Installation and Operation Manual

X-PD-BM-Ovals-eng

Part Number: 541B031AAG

January, 2009

BM Oval Series

Alarm 01 - 03:
When Alarm is displayed, please consultSection 4 for Problem solving.

Programming Setup-Level:
Configuration of the Standard Display is done at SETUP-level. SETUP-
level can be reached by pressing the PROG/ENTER key for 7 seconds; at
which time, both arrows will be displayed. In order to return to the operator
level, PROG will have to be pressed for three seconds. Alternatively, if no
keys are pressed for 2 minutes, the unit will exit SETUP automatically.
SETUP can be reached at all times and the Standard Display remains full
operational.

Password: A password may be required to enter SETUP. Without this
password access to SETUP is denied.

SETUP is divided into several main- and sub-functions. Main functions are
selected with (u) and sub-functions with (p). Each function has a unique
number, which is displayed below the word “SETUP” at the bottom of the
display. The number is a combination of two figures. The first figure
indicates the main-function and the second figure the sub-function: For
example 12 is main-function 1 and sub function 2. Furthermore, each
function is expressed with a keyword. After selecting a sub-function, the
next main function can be selected after scrolling through all “active” sub-
functions (like 1

, 11, 12, 13, 14, 1, 2, 3, 31 etc.).

Figure 3-3 Control Panel Display, Example of Function

Unit options

Function Keyword

Main Function

Sub Function

USGAL

SETUP has been divided as follows:
14

Decimals K-factor: 0 - 6

13

K-factor: 0.000010 - 9,999,999

12

Decimals: 0 - 1 - 2 -3

11

Unit: L - m3 - kg - lb - GAL - USGAL - bbl - no unit

27

Cut-off: 0.1 - 999.9 seconds

26

Calculation: per 1 - 255 pulses

25

Decimals K-factor: 0 - 6

24

K-factor: 0.000010 - 9,999,999

23

Decimals: 0 - 1 - 2 - 3(Ref: Display value)

22

Time Unit: sec - min - hour - day

21

Unit: mL - L - m3 - mg - g - kg - ton - GAL - bbl - lb - cf - rev - no unit

31

Function: total - flowrate

42

Battery Mode: operation - shelf

41

LCD Update: fast - 1sec - 3sec -15sec - 30sec - off

51

Signal: NPN - NPN_LP - Reed - Reed_LP

65

Tag number: 0000000 - 9999999

64

Password: 0000 - 9999

63

Serial No:

62

Software Version:

61

Type/Model:

Total

1

Flowrate

2

Display

3

Power Management

4

Flowmeter

5

Others

6