Brooks, Digital mfc's & mfm's – Brooks Instrument SLA5800 Series User Manual

Page 25

Advertising
background image

2-3

Brooks

®

Digital MFC's & MFM's

Section 2 Installation

Installation and Operation Manual
X-TMF-SLA5800-MFC-eng
Part Number: 541B027AAG
April, 2013

2-8 In-Line Filter

Unless an integrated (internal) filter is already installed, it is recommended
that an in-line filter be installed upstream from the mass flow controller or
meter to prevent the possibility of any foreign material entering the flow
sensor or control valve MFC. The filtering element should be replaced
periodically or ultrasonically cleaned.

Table 2-1 Recommended Filter Size

2-9 Installation

Recommended installation procedures:
a. The Brooks Digital MFC or MFM should be located in a clean,

dry atmosphere relatively free from shock and vibration.

b. Leave sufficient room for access to Self-zero function push-button.
c. Install in such a manner that permits easy removal if the instrument

requires servicing.

d. The Brooks Digital MFC or MFM can be installed in any position.

However, mounting in orientations other than the original factory
calibration(see calibration data sheet supplied with the instrument) can
result in a<±0.2% maximum full scale shift after re-zeroing.

Models

Maximum Flow Rate

Recommended Filter

SLA5850/60

100 ccm

2 micron

SLA5850/60

500 ccm

2 micron

SLA5850/60

1 to 5 lpm

10 micron

SLA5850/60

10 to 100 lpm

40 micron

SLA5851/61

10 to 30 lpm

40 micron

SLA5853/63

> 100 lpm

Consult factory

Note: Brooks provides many filter options. For those not listed here, please contact factory.