4-1-2 System Checks

The Brooks Digital Series Flowmeters and Controllers are generally used
as a component in gas handling systems, which can be complex in nature.
It can therefore be very difficult to isolate a malfunction in the system. An
inaccurately diagnosed malfunction can cause many hours of unnecessary
downtime. If possible, perform the following system checks before
removing a suspect Mass Flow Meter or Controller for bench
troubleshooting or return to the factory. (especially if the system is new):

1. Verify a low resistance common connection and that the correct power

supply voltage and signals are present of the connector of the Smart
TMF.

Remove the sensor connector from the PC Board for this procedure.

OHMMETER CONNECTION

RESULT IF ELECTRICALLY FUNCTIONAL

Open circuit on ohmmeter. If either heater (1) or

Pin 1 or 4 to meter body

sensor common (4) are shorted, an ohmmeter
reading will be obtained.

Pin 4 to Pin 2

Nominal 1100 ohms reading, depending on

Pin 4 to Pin 3

Pin 5 to Pin 1

Nominal 1000 ohm reading.

Pin 6 to Pin 7

Nominal 580 ohm reading.

temperature and ohmmeter current.

SENSOR

SCHEMATIC

PIN

NO.

FUNCTION

1

Heater

Upstream

2

Temperature

Sensor (Su)

Downstream

3

Temperature

Sensor (Sd)

4

Sensor Common

5

Heater Common

6

Thermistor

7

Thermistor

Table 4-1 Sensor Troubleshooting