Wiring the "cond" conductivity module – Burkert Type 8619 User Manual

Page 24

Advertising
background image

22

Installationandwiring

ME

RE

RE

GD

CG

FE

SE

TS

TS

1 2 3 4 5 6 7 8 9

Oxidation reduction potential

measurement electrode

Reference electrode

Temperature

sensor

Strap (not delivered)

FE = functional earth

Figure 14 :

Wiring an oxidation reduction potential sensor and a Pt100 or Pt1000 temperature sensor in a pH/ORP
module

ME

RE

RE

GD

CG

FE

SE

TS

TS

1 2 3 4 5 6 7 8 9

pH

measurement

electrode

Reference

electrode

Temperature

sensor

Oxidation reduction potential

measurement electrode

FE = functional earth

Figure 15 :

Wiring a pH sensor, an oxidation reduction potential sensor and a Pt100 or Pt1000 temperature sensor in a
pH/ORP module

7.3.6.

Wiring the "conD" conductivity module

Conductivity

sensor

C+

P+

P-

C-

GD

FE

SE

TS

TS

1 2 3 4 5 6 7 8 9

Temperature

sensor

FE = functional earth

Figure 16 :

Wiring a resistive conductivity cell with 2 electrodes and a Pt100 or Pt1000 temperature sensor in a conduc-
tivity module

Type 8619

english

Advertising