Calibrating a redox sensor – Burkert Type 8619 User Manual

Page 34

Advertising
background image

32

Adjustmentandcommissioning

8.9.4.

calibrating a redox sensor

• Modify the default calibration limits before calibrating your sensor in the menu "Parameters" -> "Mx:pH/

ORP" -> "Calibration limits" -> "Offset ORP".

• In order not to interrupt the process, activate the HOLD function in the menu "Calibration -> System ->

Hold". The Hold mode is automatically deactivated when the multiCELL restarts after a power interruption,
if the Hold mode was activated at the moment of the power cut-off.

• Before each calibration, correctly clean the electrode with a suitable product.

Calibrate the redox sensor using a 1-point procedure.

The calibration may fail:

• a possible "warning" message indicates either an error in the buffer solution or the ageing of the probe.

• a possible "error" message indicates that the probe must be replaced.

The 1-point calibration procedure is used for a quick calibration by adjusting the zero of the measurement graph
with a buffer solution with a known oxidation reduction potential.

Immerse the clean probe in the redox solution:
the multiCELL displays the measured potential
difference of the solution

Enter the potential difference of the redox
solution (indicated on the bottle).

Calibration

ORP calibration

Mx:pH/ORP

1st point

OK

F4

1st point
465.0 mV
475.0 mV

ORP Calibr. result

Offset:-55.60 mV

The multiCELL displays the calibration result.

1)

Type 8619

english