calibrating the measuring sensors

8.9.1.

calibrating a flow sensor

Enter the K factor in pulse/liter unique to the fitting used. Refer to the user manual of the fitting used.

Mx:Inputs

Calibration

DI1/2:Flow

M0:Inputs

K factor

ENTERING

8.9.2.

calibrating a conductivity sensor

Choose the reference solution used in the menu "Parameters -> Mx:Conductivity -> Buffer".

Calibrate the conductivity sensor by automatically determining its specific C constant.

Calibration

Automatic calib

Mx:Conductivity

RESULT

8.9.3.

calibrating a ph sensor

• In order not to interrupt the process, activate the HOLD function in the menu "Calibration -> System ->

Hold". The Hold mode is automatically deactivated when the multiCELL restarts after a power interruption,
if the Hold mode was activated at the moment of the power cut-off.

• Before each calibration, correctly clean the electrode with a suitable product.

Choose the buffer solution used, in the menu "Parameters -> Mx:pH/ORP -> Buffer". The multiCELL auto-
matically recognizes the pH of the buffer used.

Automatically calibrate the pH sensor:

Calibration

Mx:pH/ORP

pH auto calib.

1st point

2nd point?

Rinse

2nd point

pH Cal.Result

YES

NON

The calibration may fail:

• a possible "warning" message indicates either an error in the buffer solution or the ageing of the probe.

• a possible "error" message indicates that the probe must be replaced.