The instrument can show a 5 digit character display for flow rate values; therefore the
maximum flow rate value that can be represented on the display is 99999 (no matter the
positioning of the decimal point). The minimum is 0.0025.

The representable measure unit depends from sensor flow rate and diameter; the
permitted units are those, set with the instrument full scale value, allow to be represented
with a numerical field which the maximum value do not exceed 99999.

Example for DN 300:

-

Full scale value: 3m/s

-

PERMITTED measure unit (example): l/s (216.00); m

3

/h (777.60); m

3

/s (0.2160)

-

NOT PERMITTED measure unit (example): l/h (777600)..

Functions in the transmitter main menu are enabled by the access codes. The information
in this manual is related to all the functions available with the L2 level. All the functions
available through higher level are protected and reserved to the service.

Description of the L2 access code

(menu “11 Internal data” pos. 11.1)

‰

with code L2 = 000000 (only with this code) you can disable the L2 code request

‰

* with L2 customized (freely chosen by the user) you can configure all the functions
up to L2 security level, entering this code whenever you enter the Main menu.

* ATTENTION: take careful note of the customized code, since there is no way for the
user to retrieve or reset it if lost.

Factory preset access codes

The transmitter is delivered with the default L2 (level 2) access code:

111111