Burkert Type 8056 User Manual

Page 32

Advertising
background image

SE56

32

MENU 10 - DIAGNOSTIC

(POS. 10.1) Meter calibration

[CALIBRATION]

Enable the calibration of the meter. The activation of this function happens pressing the key Enter
during the function visualization. Will be visualized the following question: "EXECUTE?" press the key
Esc/Del/Canc to proceed. Press any other key to cancel the operation
(POS. 10.2) Autotest function enable

[SELF TEST]

Meter autotest function. This function stops the normal functions of the meter and performs a
complete test cycle on the measure input circuits and on the excitation generator. To activate this
function, after select it, push key Enter, at the question: “EXECUTE?” push the key Esc/Del/Canc.
For start autotest, or any other key for delete operation. The result of the test is shown on the
display. At the end of operation will have visualized one of visualization page. This function is
automatically performed when switching on the device.
(POS. 10.3) Flow rate simulation

[SIMULATION=ON/OFF]

Flow rate simulation enabling. With this function it is possible to generate an internal signal that
simulates the flow rate, allowing the outputs and all the connected instruments test. After enabling it,
a ‘S’ appears in the top left of the screen and the flow rate simulation can be:
‰ set: by pushing the key Enter from one of visualization pages, to set the required flow rate

(Fl.rate=) and the same key to confirm the value;

‰ finished: by pushing the key Enter from visualization pages and then by the key Esc/Del/Canc.

Advertising
This manual is related to the following products: