3 dosing quantity 8.4 selecting the dosing mode – Burkert Type 7615 User Manual

Page 12

Advertising
background image

12

Start-up

8.3

Dosing quantity

The dosing unit is set to a pumping volume of 5µl/pump stroke as
standard. This value cannot be changed.

8.4

Selecting the dosing mode

By default, the dosing unit is configured so that it doses at a fre-
quency of 5 Hz once the power supply voltage is connected. The
dosing mode can be pre-selected on the rotary switch for the elec-
tronics once the cover has been removed (see “Fig. 6: Setting the
dosing mode”
).

DANGER!

Risk of electric shock!

▶ Before reaching into the device or equipment, switch off the power

supply and secure to prevent reactivation!

▶ Only adequately trained personnel may reach into the device!

Rotary switch for pre-selecting the

dosing mode

Fig. 6: Setting the dosing mode

Making adjustments using the rotary switch for the electronics:

0 Impulse mode

>120 ms

Unit doses precisely 5 µl
(= 1 stroke) and then stops.
Power supply voltage must be at
least 120 ms for this purpose. If the
power supply is then interrupted
for > 2 ms and is subsequently
reconnected, a stroke is initiated, etc.

1 40 Hz frequency mode Unit is in frequency mode and

doses at the selected frequency
once the power supply voltage is
connected.

2 25 Hz frequency mode
3 10 Hz frequency mode
4 5 Hz frequency mode

(standard)

5 5 Hz service mode;

unit stops after
100 cycles

For servicing purposes: Following
connection of the power supply
voltage, the unit operates at a fre-
quency of 5 Hz for precisely
100 pumping cycles and then
stops.
The supply rate can be checked
using this function.

english

Type 7615