Fluid Components International FLT Series User Manual

Page 62

Advertising
background image

FLUID COMPONENTS INTL

APPENDIX D - TEMP COMP

FLTÔ Series FlexSwitchÔ

D - 4

Doc. No. 06EN003286 Rev. -

6.

Stop the process media flow and make sure that the media is at no flow and then let the instrument stabilize
for fifteen minutes.

7.

Record the resistance values of the sensing elements and calculate the resistance differential (DR). If DR does
not exceed the maximum DR of 280 ohms then proceed with the calibration. If the DR is above 280 ohms use
the next lower heater wattage setting and let the instrument stabilize. Recheck the DR

8.

Start the process media flowing at the desired switch point velocity and at the low temperature, let the
instrument stabilize for fifteen minutes.

9.

Record the resistance values of the active and reference RTD's at the low temperature.

10. Raise the temperature of the process media to the maximum expected temperature. With the instrument power

on, let the instrument stabilize for fifteen minutes. The difference between the low and the high temperature
should not exceed 100°F.

11. Record the resistance values of the active and reference RTD's for the high temperature.

12. Calculate the Temp Comp factor with the formula shown below.

TEMP

COMP =

D

R Low Temperature - DR High Temperature .

FACTOR (R Reference High Temperature) - (R Reference Low Temperature)

The Temp Comp factor is not to exceed ±0.041.

13. If the Temp Comp factor is within tolerance, turn off the power to the instrument and stop the process media if

needed. Disconnect the DMM's and the power supply from the instrument. Reconnect the sensing element
wires to the control circuit socket and reinstall the socket in the enclosure if it was previously removed. Do
not pinch the wires between the socket and the enclosure.

14. Look up the resistance values to adjust potentiometers R5 and R8 in the Temp Comp Factor table (Table D-2).

Follow the procedure in the Restoring Temp Comp adjustment section using the values found in the table
below.

15. If the calculated Temp Comp factor exceeds the allowable tolerance by a small amount (±0.01), using the

maximum Temp Comp factor may make the instrument perform satisfactorily. However, if the factor is out of
tolerance by more than ±0.01 then it will be necessary to repeat the calibration to verify the result. Continue
with the adjustment procedure if the second result is within tolerance.

Advertising