Fluid Components International FLT Series User Manual

Page 65

Advertising
background image

FLUID COMPONENTS INTL

APPENDIX E

FLT

Ô Series FlexSwitchÔ

E - 2

Doc. No. 06EN003246 Rev. B

INSTALLATION CONFORMITY CRITERIA

Grounding

All enclosures must be grounded to earth ground through a path of less than 1 ohm.

Interconnecting Cables

All interconnecting cables between the FlexSwitch local enclosure, remote enclosure, power source and monitoring
device shall be enclosed in metal conduit. AC power input cabling shall be enclosed separately in conduit for
entrance into the FlexSwitch and is not to be combined with switch or monitor output cabling.

Standard ESD Precautions

Use standard ESD precautions when opening an instrument enclosure or handling the FlexSwitch. FCI recommends
the use of the following precautions: Use a wrist band or heel strap with a 1 megohm resistor connected to ground.
If the instrument is in a shop setting there should be static conductive mats on the work table and floor with a 1
megohm resistor connected to ground. Connect the instrument to ground. Apply antistatic agents to hand tools to
be used on the instrument. Keep high static producing items away from the instrument such as non-ESD approved
plastic, tape and packing foam.

The above precautions are minimum requirements to be used. The complete use of ESD precautions can be found in
the U.S. Department Of Defense Handbook 263.

Location of CE mark documentation (European Location)

The technical documentation file part A resides at Fluid Components Intl, European Service Center, Persephonestraat
3-01 5047 TTTilburg - The Netherlands - Phone 31-13-5159989 - Fax 31-13-5799036

Location of CE mark documentation (Manufacturer Location)

The technical documentation file part B resides at the Configuration Management department of Fluid Components
Intl, 1755 La Costa Meadows Dr. San Marcos, Ca 92069 USA.

Advertising