Error icon (blinks) problem possible causes – Hach-Lange SI794 D tr User Manual User Manual

Page 83

Advertising
background image

Alarm contact

Red LED

out

1 (22 mA)

Out 2 (22 mA)

Current output 1

Current below 0 (3.8) mA

Error

Icon

(blinks)

Problem

Possible causes

ERR 11

x

x

x

Temperature probe

Open or short circuit

Temperature range exceeded

ERR 03

x

x

x

x

Current output 1

Current above 20.5 mA

ERR 12

x

x

x

Current output 1

Current span too small / too large

ERR 13

x

x

x

Current output 2

Current below 0 (3.8) mA

ERR 21

x

x

x

Current output 2

Current above 20.5 mA

ERR 22

x

x

x

Current output 2

Current span too small / too large

ERR 23

x

x

x

Rinsing probe

Communication error

ERR 41

x

x

x

Sensocheck

Sensor: connecting cable defective

ERR 33

x

x

x

• Zero error, Sensoface active, see Pg 86

• Slope error, Sensoface active, see Pg 86

x

• Response time exceeded, Sensoface active, see Pg 86

• Cal interval expired, Sensoface active, see Pg 86

83