Hach-Lange SI794 D tr User Manual User Manual

Page 91

Advertising
background image

Temperature input

NTC 22 kohms / NTC 30 kohms*

2-wire connection, adjustable

Measuring range

-20.0 ... +150.0 °C / -4 ... +302 °F

Adjustment range

10 K

Resolution

0.1 °C / 1 °F

Measurement error

1,2,3

< 0.5 K (<1 K at > 100°C)

HOLD input

Galvanically separated (OPTO coupler)

Function

Switches transmitter to Hold mode

Switching voltage

Inactive 0 ... 2 V (AC/DC)

Active 10 ... 30 V (AC/DC)

CONTROL input

Galvanically separated (OPTO coupler)

Function

Control input for automatic cleaning system

Switching voltage

Inactive 0 ... 2 V (AC/DC)

Active 10 ... 30 V (AC/DC)

Output 1

0/4 to 20 mA, max. 10 V, floating

(galv. connected to output 2)

Process variable *

O

2

saturation/O

2

concentration

Overrange *

22 mA in the case of error messages

Output filter *

Low-pass, filter time interval 0 ... 120 s

Meas. error

1

< 0.3 % current value + 0.05 mA

Start/end of scale

Configurable within selected range

Span allowed

5 ... 500 % / 0.5 ... 50 mg/L (ppm)

Output 2

0/4 to 20 mA, max. 10 V, floating

(galv. connected to output 1)

Measured variable

Temperature

Overrange *

22 mA in case of temp error messages

Output filter *

Low-pass, filter time interval 0 ... 120 s

Measurement error

1

< 0.3 % current value + 0.05 mA

Start/end of scale *

-20 ... +150 °C / -4 ... +302 °F

Span allowed

20 ... 170 K (68 ... 338 °F)

91