Communication menu – Hach-Lange ORBISPHERE K1100_M1100 Operator Manual User Manual

Page 82

Advertising
background image

80

Communication Menu

Commands

The “Command Output Buffer” is formatted as follows:

The “Command Input Buffer” is located just after the measurement data and is formatted as
follows:

The following commands are available:

Change product

Activate sensor (valid for EC sensors only)

Change product command - output

Change product command - input

Name

Type

Size

Offset

Output command toggle (OCT)

Output byte

8 bits

0

Output command ID (OCI)

Output byte

8 bits

1

Output command data byte 1 (OCD1)

Output byte

8 bits

2

Output command data byte 2 (OCD2)

Output byte

8 bits

3

Output command data byte 3 (OCD3)

Output byte

8 bits

4

Output command data byte 4 (OCD4)

Output byte

8 bits

5

Name

Type

Size

Offset

Input command toggle (ICT)

Input byte

8 bits

74

Input command status (ICS)

Input byte

8 bits

75

Input command data byte 1 (ICD1)

Input byte

8 bits

76

Input command data byte 2 (ICD2)

Input byte

8 bits

77

Input command data byte 3 (ICD3)

Input byte

8 bits

78

Input command data byte 4 (ICD4)

Input byte

8 bits

79

Name

Value

Comment

OCT

1-2

OCI

1

OCD1

0

Channel number:
0 = Channel 1

OCD2

0-99

Product number

OCD3

0-1

Erase measurement files:
0 = Never erase the measurement files.
1 = Erase measurement file if necessary (e.g. gas unit changes)

OCD4

Not

used

Name

Value

Comment

ICT

1-2

ICS

0-3

0 = OK
1 = Unknown command ID
2 = Invalid parameter (e.g. invalid channel no or product number)
3 = Execution failure

ICD1

Not used

ICD2

Not used

ICD3

Not used

ICD4

Not

used

Advertising