3 usb-a port (host), Communication menu – Hach-Lange ORBISPHERE K1100_M1100 Operator Manual User Manual

Page 83

Advertising
background image

81

Communication Menu

Activate sensor command - output

Activate sensor command - input

9.3 USB-A port (host)

This option allows the export or import of data from an external mass storage device. The
device must first be connected to the instrument through the USB-A port.

Select one of the two import options (product list or access table) to import data from the
storage device. This is useful for transferring these files to additional instruments without the
need of having to re-enter the data individually on each instrument.

Note:

The imported data will override any current settings on the instrument.

Select the export option to export data from the instrument to the storage device. For
information regarding the uploaded files, refer to

Uploaded Files on page 85

.

For both import and export options, the progress bar is updated to give an indication of the
progress of the selected option.

Name

Value

Comment

OCT

1-2

OCI

2

OCD1

0

Channel number:
0 = Channel 1

OCD2

0-1

Sensor activation:
0 = Deactivate the EC sensor
1 = Activate the EC sensor

OCD3

Not used

OCD4

Not

used

Name

Value

Comment

ICT

1-2

ICS

0-3

0 = OK
1 = Unknown command ID
2 = Invalid parameter (e.g. invalid channel no)
3 = Execution failure

ICD1

Not used

ICD2

Not used

ICD3

Not used

ICD4

Not

used

Advertising