Using the Power Sensor

Making Measurements

MA24106A UG

3-13

Table 3-1, describes the number of averages needed to attain a certain noise level for a particular power level
measurement with the Low Aperture Time mode setting.

Table 3-2, describes the number of averages needed to attain a certain noise level for a particular power level
measurement with the High Aperture Time mode setting.

Table 3-1.

MA24106A Averaging Table (Low Aperture Time, Default Mode)

Input Power

(dBm)

Input Power

(mW)

Number of

Averages

Needed for

<

±

0.20 dB

Noise

Number of

Averages

Needed for

<

±

0.15 dB

Noise

Number of

Averages

Needed for

<

±

0.10 dB

Noise

Number of

Averages

Needed for

<

±

0.05 dB

Noise

Number of

Averages

Needed for

<

±

0.01 dB

Noise

20

100

1

1

1

1

1

15

31.6

1

1

1

1

1

10

10.0

1

1

1

1

1

5

3.16

1

1

1

1

2

0

1.00

1

1

1

4

16

-5

0.316

1

1

1

20

78

-10

0.100

1

1

1

1

1

-15

0.0316

1

1

1

1

1

-20

0.0100

1

1

1

1

7

-25

0.00316

1

1

1

3

61

-30

0.00100

2

3

7

25

-35

0.000316

16

28

62

245

-40

0.000100

158

Table 3-2.

MA24106A Averaging Table (High Aperture Time)

Input Power

(dBm)

Input Power

(mW)

Number of

Averages

Needed for

<

±

0.20 dB

Noise

Number of

Averages

Needed for

<

±

0.15 dB

Noise

Number of

Averages

Needed for

<

±

0.10 dB

Noise

Number of

Averages

Needed for

<

±

0.05 dB

Noise

Number of

Averages

Needed for

<

±

0.01 dB

Noise

20

100

1

1

1

1

1

15

31.6

1

1

1

1

1

10

10.0

1

1

1

1

1

5

3.16

1

1

1

1

1

0

1.00

1

1

1

1

1

-5

0.316

1

1

1

2

5

-10

0.100

1

1

1

1

1

-15

0.0316

1

1

1

1

1

-20

0.0100

1

1

1

1

1

-25

0.00316

1

1

1

1

4

-30

0.00100

1

1

1

2

38

-35

0.000316

1

2

4

16

-40

0.000100

10

18

39

153