Remote Operation

Remote Operation Command Details

MA24106A UG

4-3

FREQ?

Description:

Gets the current cal factor frequency value from the sensor.

Syntax:

FREQ? + LF

Return Value:

Current cal factor frequency in GHz

FREQ

Description:

Sets the current cal factor frequency value for the sensor.

Syntax:

FREQ fghz + LF

Return Value:

OK or ERR

Remarks:

“fghz” is the cal factor frequency value in GHz. “fghz” must be between 0.05 GHz to
6 GHz. Available resolution is 3 digits after the decimal point.

ZERO

Description:

Zeros the power sensor.

Syntax:

ZERO + LF

Return Value:

OK if successful and ERR if zeroing failed.

Remarks:

In case of zero failure, the STATUS? command can be used to retrieve more detail about
the error. See the STATUS? command for more detail. It takes 19 seconds to zero a
sensor, please wait at least this long to get a response from the sensor.

TMP?

Description:

Gets the current temperature reading from the sensor.

Syntax:

TMP? + LF

Return Value:

Current temperature in degrees C.

STATUS?

Description:

Get error status byte from the sensor.

Syntax:

STATUS? + LF

Return Value:

Error status byte

Remarks:

Error status byte information:

Status.b0 -> ZERO_TEMP_ERROR (Temperature changed more than

allowable limit after zeroing sensor)

Status.b1 -> Not Used

Status.b2 -> ADC_CH2_OR (Temperature over range)

Status.b3 -> ADC_CH3_OR (Detector A over ranged)

Status.b4 -> ZERO_ERROR_DET_A

Status.b5 -> ZERO_ERROR_DET_B

Status.b6 -> TEMP_ERROR (Temperature beyond operating range)

Status.b7 -> Not Used