Linearity Test

Sensor Operational Tests

5-6

MA24106A UG

n. Increase averaging on the MA24106A by clicking the Averages button, enter “16” and then click

Apply

.

2. Apply the appropriate Cal factor to the reference sensor per the manufacturer’s procedure.
3. Apply the appropriate Cal factor to the MA24106A as follows:

Click the Frequency button on the Power Meter application, and then enter the frequency of the
measurement in GHz.

4. Turn Off the synthesizer’s RF output and perform a low-level Zero of both the Reference sensor and the

MA24106A.

5. Turn On the synthesizer’s RF output.
6. Record data for the first 20 dB range as follows:

a. Record the power reading by the reference meter in the appropriate space in Table 5-4.
b. Record the power reading by the MA24106A in the appropriate space in Table 5-4.
c. Set the synthesizer power to +15 dBm.
d. Record the reference meter and the MA24106A power sensor readings in the appropriate places in

Table 5-4.

e. Repeat the measurement for synthesizer output levels of +10, +5, and 0 dBm.

Figure 5-2.

Linearity Test Setup

Note: The MA24106A power measured at 0 dBm will be used in step 7e, below.