Anritsu 0701012 User Manual

Page 50

Advertising
background image

Linearity Test

Sensor Operational Tests

5-10

MA24106A UG

Table 5-7.

Measurement Results (6 GHz)

Row

#

Synthesizer

Power Level

Setting

(dBm)

Attenuation

in Reference

Arm
(dB)

A

Reference

Power

Measurement

(dBm)

B

= (A6–A5)

Correction

(dB)

C

= (A+B)

Corrected

Reference Power

Measurement

(dB)

Attenuation

in Test Arm

(dB)

D

MA24106A

Measurement

(dBm)

E

= (C–D)

Difference

Calculation

(dB)

1

+20

10

0

2

+15

10

0

3

+10

10

0

4

+5

10

0

5

0

10

0

6

adjust per

step 7e

0

0

10

7

+5

0

0

10

8

0

0

0

10

9

–5

0

0

10

10

–10

0

0

10

11

Max:

12

Min:

13

Delta (E11 – E12):

Advertising
This manual is related to the following products: