Uncertainty of a Measurement

Using the Power Sensor

3-18

MA24106A UG

Uncertainty Example

An example of a measurement uncertainty calculation (for Low Aperture Time mode) is detailed for the
MA24106A in Table 3-4 below. The MA24106A is used to measure the power of a 3 GHz, +12.0 dBm and
–35 dBm CW signal from a signal source with 1.5:1 VSWR.

Table 3-4.

Measurement Uncertainty Example

Uncertainty

Term

Uncertainty

Specification

at +12 dBm

(%)

Uncertainty

Specification

at –35 dBm

(%)

Probability

Distribution

Divisor

Adjusted

Uncertainty

at +12 dBm

(%)

Adjusted

Uncertainty

at –35 dBm

(%)

Sensor Linearity

(<+18 dBm)

3.0

3.0

Rectangular

1.8

1.8

Noise

0.0

0.8

Normal at 2

σ

2

0.0

0.4

Zero Set

0.0

3.2

Rectangular

0.0

1.8

Zero Drift

0.0

0.9

Normal at 2

σ

2

0.0

0.5

Calibration Factor

Uncertainty

1.4

1.4

Normal at 2

σ

2

0.7

0.7

Mismatch

Uncertainty

4.0

4.0

Rectangular

2.3

2.3

Combined

Uncertainty

(RSS), Room

Temperature

3.0

3.6

Expanded

Uncertainty with

K=2, Room

Temperature

6.0

7.1

Temperature

Compensation

1.4

1.4

Rectangular

0.8

0.8

Combined

Uncertainty

(RSS, 0 to 50 °C)

3.1

3.6

Expanded

Uncertainty

with K=2

(RSS, 0 to 50 °C)

6.2

7.3

Table 3-5.

Noise Measurement Uncertainty Calculations

Noise Calculations at 12 dBm (16 mW):

Noise 400 nW/16 mW = 0.0%

Zero Set 1700 nW/16 mW = 0.0%

Zero Drift 500 nW/16 mW = 0.0%

Noise Calculations at –35 dBm (316 nW):

Noise 2.5 nW/316 nW = 0.8%

Zero Set 10 nW/316 nW = 3.2%

Zero Drift 3 nW/316 nW = 0.9%