Syringe seal – Luminex 100 IS User Manual, Version 2.1 User Manual

Page 142

Advertising
background image

100 IS User Manual Version 2.1

x

MAP Technology

8 - 6

PN 89-00002-00-070 Rev. A

Syringe Seal

‹

To replace the syringe plunger seal:

1. Open the center-most door on the front of the analyzer.

2. Locate the syringe (a glass cylinder with a metal rod).

Figure 103. Syringe and Syringe Seal

3. Unscrew the knob on the syringe arm (at the bottom of the

syringe), and forcefully push the syringe arm down.

4. Unscrew the syringe from the top of its housing.

5. Pull the plunger out of the syringe.

6. Remove and replace the plunger seal.

7. Return the plunger to the syringe.

8. Screw the syringe back into its housing.

9. Return the syringe arm to its original position.

Warning:

Turn the unit OFF and unplug the power cord before

replacing the syringe plunger! The syringe arm does not deactivate
when changing the plunger; injury could result if the system is not
unplugged.

Syringe seal

Syringe