System overview, Recommended additional equipment – Luminex 100 IS User Manual, Version 2.1 User Manual

Page 49

Advertising
background image

x

MAP Technology

The System

PN 89-00002-00-070 Rev. A

5 - 7

Power cords: Power cords specific to the country of use are
included in the system as required

Power: 115-230 V~, 5.6 Amps, 50-60 Hz

Recommended Additional
Equipment

Uninterruptible Power

Supply (UPS)

Luminex highly recommends using an uninterruptible power supply
(UPS) to protect your system from a power outage. Choose a supply
that can provide 1050 Watts for at least 45 minutes. The UPS should
be UL listed, CSA certified, CE marked when used internationally.

Surge Protector

If you do not use a UPS, as a minimum, use a surge protector.
Choose a protector that meets your needs. Factors to consider include
electrical environment, endurance, suppressed voltage rating, and
method of protection. It should have six outlets, rated at least 1500
Watts, and be UL listed, CSA certified, CE marked for nondomestic
use when used internationally

Printer

Printer, HP LaserJet 2100xi

Bar Code Labels

Use the Code 128 barcode label type when scanning bar code labels
into the system as patient identities.

Vortex

VWR product number 58816-121: Speed range 0 - 3200 rpm or
equivalent

Bath Sonicator

Parmer product number 08849-00: Operating frequency 55 kHz
or equivalent

System Overview

The system consists of three subsystems: electronic, fluidic, and
optical. The following section describes the user-accessible
components of each subsystem.

Advertising