Luminex 100 analyzer ventilation filter – Luminex 100 IS User Manual, Version 2.1 User Manual

Page 143

Advertising
background image

x

MAP Technology

Maintenance and Cleaning

PN 89-00002-00-070 Rev. A

8 - 7

10. Hand-tighten the screw on the syringe arm.

11. Plug in the power cord and turn the analyzer on.

12. Prime the system twice, watching for any leaks in the syringe

area.

13. When the prime finishes, close the center door.

Luminex 100 Analyzer
Ventilation Filter

Clean the filter only when soiled.

‹

To check the Luminex 100 analyzer ventilation filter, follow

these steps:

1. Disconnect the instrument from ac power by turning off the

power switch on the rear of the instrument, then unplugging the
instrument power cord from the wall source.

2. On the lower right side of the Luminex 100 analyzer, firmly push

the filter to the left.

3. Remove the filter from the left side of the Luminex 100 analyzer.

Figure 104. Luminex 100 Analyzer Ventilation Filter

4. Clean the filter with a vacuum or with distilled water. Stand the

filter upright to air dry.

5. Re-install it with the arrows facing up. The filter should click

into place.

6. Plug in and power on the Luminex 100 analyzer.

Luminex 100 analyzer ventilation filter (on bottom)