Discontinued – Max Machinery 164 SERIES SIGNAL CONDITIONER_CONVERTER User Manual

Page 7

Advertising
background image

164-110-350 Max Machinery, Inc.

Page 7

Example:

213 Flow Meter

284 Transmitter (5V Square Wave Output)

Set S10-2 to “ON”,(5 volt position).

Maximum Flow Rate = 1000 cc/min

K-Factor - 115 pulses/cc

Maximum Frequency = 1000 x 115 = 1917 (1.917 KHz)

60

Set S10-5 to “ON”; S9-2 to 3 KHz.

Converter Volts = 1.92 KHz x 10 = 6.4 V

3.0 KHz

Example:

241 Flow Meter

289 Transmitter (15V Square Wave Output)

Set S10-1 to “ON”.

Maximum Flow Rate = 30 L/min

K-Factor = 403 pulses/liter

Maximum Frequency = 30 x 403 = 201.5 Hz

60

Set S10-5 to “ON”; S9-4 to 300 Hz.

Converter Volts = 201.5 Hz x 10 = 6.72 V

300 Hz

Hysteresis Adjust (Sine Wave Inputs only): Set the system flow rate to about one third of the
minimum operational level. Attach an oscilloscope to TP2 located between S9 and S10.
Adjust the pot until the pulses on the display start to drop off to zero. Any flow lower than
this level will not produce a display or output.

FREQUENCY TO DC CONVERTER CALIBRATION